Kawasaki Ninja Service Manual: Pressure Testing

Kawasaki Ninja Service Manual / Cooling System / Coolant / Pressure Testing

Remove the radiator cap in two steps. First turn the cap counterclockwise to the first stop. Then push and turn it further in the same direction and remove the cap.

Pressure Testing


NOTE

Wet the cap sealing surfaces with water or coolant to prevent pressure leaks.

Pressure Testing


NOTICE

During pressure testing, do not exceed the pressure for which the system is designed. The maximum pressure is 142 kPa (1.45 kgf/cm², 21 psi).

If the pressure holds steady, the system is all right.

If the pressure drops and no external source is found, check for internal leaks. Droplets in the engine oil indicate internal leakage. Check the cylinder head gasket and the water pump.

Remove the pressure tester, replenish the coolant, and install the radiator cap.

Coolant Deterioration Inspection
Remove the right lower fairing (see Lower Fairing Removal in the Frame chapter). Visually inspect the coolant in the reserve tank [A]. If whitish cotton-like wafts are observed, aluminum par ...

Cooling System Flushing
Over a period of time, the cooling system accumulates rust, scale, and lime in the water jacket and radiator. When this accumulation is suspected or observed, flush the cooling system. If this accu ...

Other materials:

Battery Charging
Remove the battery from the motorcycle (see Battery Removal). Attach the leads from the charger and charge the battery at a rate (amperage x hours) that is indicated on the battery. If it is not possible to read the rate, charge the battery at an amperage that is about 1/10th of the ...

System Components
1. KIBS ECU With input from the front and rear wheel sensors and front caliper hydraulic pressure, the KIBS ECU analyzes various conditions on the motorcycle, and with additional information from the FI ECU (engine speed, throttle position, gear position, and clutch actuation) precisely control ...

Check 3-7 Stop Watch Inspection
Connect the leads in the same circuit as Check 3-2. By pushing the upper button each time to set the stop watch mode. Connect the insulated auxiliary lead processed insulation to the terminal [13] as shown in the figure, then stop watch start to count. While count the stop watch, conn ...