Right-hand End Cover and Negative Brushes [A] Terminal Bolt and Positive Brushes [B]
Special Tool - Hand Tester: 57001-1394
If there is not close to zero ohms, the brush lead has an open. Replace the brush plate assy.

Armature Inspection
Right-hand End Cover InspectionCheck 3-7 Stop Watch Inspection
Connect the leads in the same circuit as Check 3-2.
By pushing the upper button each time to set the stop
watch mode.
Connect the insulated auxiliary lead processed insulation
to the terminal [13] as shown in the figure, then stop watch
start to count.
While count the stop watch, conn ...
Instrument display brightness control
The brightness of the tachometer
LED segments and illumination of the
multifunction meter are controlled automatically
depending on the ambient
brightness.
A. Tachometer
B. Ambient Brightness Sensor
C. Multifunction Meter
NOTE
Be careful not to cover the ambient
brightness sensor on th ...
KIBS Servicing Precautions
There are a number of important precautions that should
be followed servicing the KIBS.
This KIBS is designed to be used with a 12 V sealed battery
as its power source. Do not use any other battery
except for a 12 V sealed battery as a power source.
Do not reverse the battery cable connec ...